Open circuit and rebreathers compared

How gas use, oxygen and decompression differ between open circuit, semi-closed and closed circuit.

10 min read

The guide to the rebreather showed how a rebreather keeps your breath in a loop and adds only the oxygen you use. This guide puts it side by side with open circuit on the same dives. It compares 3 things: how much gas each uses, how much oxygen each gives you, and how much decompression each needs.

On open circuit (OC) you breathe from a cylinder through a regulator, and every breath you exhale leaves as bubbles. On a closed-circuit rebreather (CCR) your exhaled gas stays in a loop. A scrubber removes the carbon dioxide, and the unit adds oxygen to replace what your body used.

All the plans in this guide were planned with DiveLogic, with the same decompression model and settings for both.

Gas use

On open circuit each breath comes out of the regulator at the ambient pressure, the total pressure around you. It is measured in bar: 1 bar at the surface, plus 1 bar for every 10 m. So at 45 m it is 5.5 bar. By Boyle’s law, a breath there holds 5.5 times as much gas as the same breath at the surface, so you use gas 5.5 times as fast.

Your RMV (respiratory minute volume, also called SAC rate) is the gas you breathe in a minute at the surface. On open circuit, your gas use per minute at any depth is your RMV scaled up by the pressure. This is the number that sizes your cylinders.

A rebreather adds only the oxygen your body uses. Your body uses the same amount of oxygen per minute at any depth. It depends on how hard you work, not on the pressure, so there is no PP in it. The rest of the loop gas is breathed again.

This changes what matters for gas use. On open circuit, every extra breath costs gas, so a fast breather empties a cylinder sooner. On a rebreather, breathing faster or slower changes little on its own. What uses up the oxygen is the work your body does: a calm diver uses less, and a diver swimming hard uses more, at any depth.

The loop also needs diluent, a second gas that keeps the loop full. As you descend, the loop gas is squeezed and the unit adds diluent. As you ascend, the gas expands and some is let out. So diluent use depends on depth changes and events such as clearing your mask, not on how long you breathe.

  • Open circuit
  • Rebreather O₂, working
  • Rebreather O₂, at rest
Open circuit at a surface breathing rate of 15 L/min. Rebreather oxygen at DiveLogic’s planning rates of 3 L/min while working and 1 L/min at rest, the same at every depth. Filling a 5 L loop with diluent on the descent to 45 m takes 23 L in total.

Planned with DiveLogicWork out the gas for a dive

DiveLogic plans rebreather oxygen at 3 L/min while working and 1 L/min at rest. These are planning figures, set higher than most divers use, because running out of oxygen ends the dive. You can work out open-circuit gas for any dive with the gas needed tool.

Oxygen pressure

The PO₂ (partial pressure of oxygen, also written ppO₂) is the pressure of the oxygen alone in the gas you breathe. It is the FO₂, the fraction of oxygen in the mix, times the ambient pressure (Dalton’s law).

On open circuit the FO₂ is fixed, so the PO₂ rises with depth. Air gives a PO₂ of 0.21 bar at the surface and 0.21 × 5.5 = 1.16 bar at 45 m. So each open-circuit mix has a MOD (maximum operating depth): the deepest point at which its PO₂ stays within your chosen limit. To breathe more oxygen on the ascent, you must carry a deco gas in a stage cylinder and make a gas switch to it.

An electronic CCR holds a fixed PO₂ at any depth, called the setpoint. Think of it like a thermostat: when the PO₂ drops below the setting, the unit adds oxygen, as a heater comes on when a room cools. A common working setpoint is 1.3 bar. Many divers descend on a lower setpoint, such as 0.7 bar, and switch to 1.3 bar at the bottom. A high setpoint cannot be held in the shallows (see the next paragraph), and the loop PO₂ rises for a moment as the gas is squeezed on the way down.

Near the surface the loop cannot hold a setpoint higher than the pressure around you. At 0 m the most it can hold is about 1 bar. So the loop PO₂ is the lower of the setpoint and the ambient pressure.

The figures in this guide are for closed circuit only. A semi-closed rebreather’s loop oxygen changes with how hard you work.

  • Air
  • EAN32
  • Setpoint 1.3
  • Descent setpoint 0.7
On open circuit the ppO₂ rises with depth. The rebreather holds 0.7 bar on the descent and 1.3 bar from the bottom to the surface. Shallower than 3 m it cannot hold 1.3 bar, because the ambient pressure is lower than that.

Planned with DiveLogicSet rebreather setpoints

Decompression

Inert gas is the part of your gas that the body does not use: nitrogen and helium. Your tissues take it up, called on-gassing, at a rate set by the inert gas pressure you breathe (see tissues and half-times). Oxygen and inert gas share the ambient pressure, like 2 people sharing a fixed-size cake: the bigger the oxygen slice, the smaller the inert gas slice. So a higher PO₂ means a lower inert gas pressure.

A rebreather holds a high PO₂ at every depth. So on most dives your tissues take up less inert gas and you have a smaller decompression obligation: fewer and shorter decompression stops on the way up. The tissue model is the same Bühlmann model for both. Only the inert gas pressure fed into it changes.

On open circuit, the inert gas pressure is the ambient pressure times the inert part of the mix (FN₂ plus FHe). On a CCR, it is the ambient pressure minus the setpoint. Both first take off a small amount, about 0.06 bar, for water vapour in the lungs. On the loop, the inert gas is split between nitrogen and helium in the same proportions as in the diluent.

At 45 m on 21/35, open circuit gives about 4.30 bar of inert gas. A 1.3 setpoint gives about 4.14 bar. Both plans count each 10 m of sea water as 1 bar.

The comparison below uses one dive: 45 m for 30 minutes. 21/35 is trimix with 21% oxygen and 35% helium; it is the open-circuit bottom gas (the gas for the deepest part of the dive) and the rebreather diluent. Both plans use gradient factors 50/80, GF low 50 and GF high 80 (see gradient factors), which set how close to its limit the model lets your tissues go.

  • Open circuit 21/35
  • EAN50
  • Rebreather, setpoint 1.3
Open circuit: 45 m for 30 min on 21/35, EAN50 from 21 m, GF 50/80. Runtime 75 min.

Planned with DiveLogicPlan this dive

The open-circuit plan

The open-circuit diver breathes 21/35 on the bottom. At 21 m they switch to EAN50, nitrox with 50% oxygen, for the decompression stops.

On the bottom, 21/35 gives a PO₂ of 1.16. From the bottom to 21 m the diver is still on 21/35, so the PO₂ falls on the way up. The TTS (time to surface) from the end of the bottom time is 43 minutes.

The same dive on a rebreather

The rebreather diver holds a PO₂ of 1.3 from the bottom to the surface. On the bottom and up to 21 m, they breathe less inert gas than the open-circuit diver. After the switch to EAN50, the open-circuit diver breathes more oxygen: a PO₂ of 1.56 at 21 m.

The rebreather plan has a shallower first stop and has a TTS of 34 minutes. That is 9 minutes less. On longer or deeper dives the difference is usually larger.

Without a decompression gas

The dashed line is the open-circuit plan with no EAN50: the diver breathes 21/35 all the way up. The TTS is now 92 minutes.

On open circuit you cut the inert gas on the ascent only by carrying extra cylinders. A rebreather gives you a high PO₂ all the way up without changing gas.

Oxygen exposure

A high PO₂ all dive also means more oxygen exposure. Divers track 2 numbers. The CNS clock is how much of the NOAA limit for CNS oxygen toxicity (oxygen fits, or seizures) you have used, as a percentage. OTU (oxygen tolerance units) measures pulmonary oxygen toxicity, the load on your lungs and whole body.

On this dive the rebreather plan ends at 36% CNS and 95 OTU. The open-circuit plan with EAN50 ends at 27% and 70 OTU. On long rebreather dives the CNS clock, not decompression, is often what limits the time you can stay. The guides to CNS oxygen toxicity and pulmonary oxygen toxicity cover both numbers.

Bailout

A rebreather can fail: the loop can flood, a sensor or valve can fail, or the scrubber can stop working. The standard response is to bail out: leave the loop and breathe open circuit.

From then on you use gas at the open-circuit rate, from the depth of the failure and through every decompression stop. So rebreather divers carry open-circuit bailout cylinders big enough for the worst point of the dive. That is usually the end of the bottom time at the deepest point, where the open-circuit ascent is longest and each breath takes the most gas. DiveLogic plans these open-circuit ascents as bailout scenarios.

Try it

Choose the depth, the bottom time, the mix and the setpoint. The figure plans the dive on open circuit with EAN50 and on a rebreather, and the table compares them. Deeper and longer dives show a bigger difference. A higher setpoint shortens the rebreather plan and raises its oxygen exposure.

  • Open circuit 21/35
  • EAN50
  • Rebreather, setpoint 1.3
Open circuitRebreather
Runtime75 min66 min
Time to surface43 min34 min
Gas planned21/35 2,775 LEAN50 823 LOxygen 160 LDiluent 120 L
CNS27%36%
OTU7096
Both plans use GF 50/80. Open circuit carries EAN50 from 21 m at a breathing rate of 15 L/min on the bottom and 12 L/min on the stops. Rebreather oxygen is planned at 3 L/min working and 1 L/min on the stops; diluent covers the descent, 2 mask clears, 1 loop flush and the wing.
Bottom gas and diluent
Setpoint

Planned with DiveLogicPlan a CCR dive

What to remember

  1. Open-circuit gas use rises with the pressure around you: at 45 m you use about 5.5 times as much gas per minute as at the surface.
  2. A closed-circuit rebreather uses only the oxygen your body uses, which does not depend on depth. It also uses some diluent when the depth changes.
  3. Open circuit has a fixed oxygen fraction, so its PO₂ rises with depth. An electronic CCR holds a fixed PO₂, the setpoint.
  4. A high PO₂ at every depth means less inert gas, so a rebreather dive usually needs less decompression than the same dive on open circuit.
  5. The same high PO₂ increases CNS and OTU exposure, and a rebreather diver still carries open-circuit bailout gas.

Rebreather plans in DiveLogic

Turn on CCR mode in the planner, choose a diluent and the descent, working and decompression setpoints, and plan the dive. The review shows the oxygen and diluent the dive needs, the CNS and OTU, and the open-circuit bailout scenarios.

Sources

  1. Naval Sea Systems Command (2016). U.S. Navy Diving Manual, Revision 7. SS521-AG-PRO-010.
  2. Vann R. D., Denoble P. J., Pollock N. W. (eds) (2014). Rebreather Forum 3 Proceedings. AAUS, DAN and PADI, Durham NC.
  3. Bühlmann A. A., Völlm E. B., Nussberger P. (2002). Tauchmedizin, 5th edition. Springer, Berlin.
  4. Hamilton R. W. (1989). Tolerating exposure to high oxygen levels: Repex and other methods. Marine Technology Society Journal 23(4): 19-25.